• • • • Bruker EMX basic operation System Startup • Sign in in the log book. • Make sure the chilled water is on. • Turn on the heat exchanger. • Turn on the power supply (Do Not Turn On The Magnet!) • Turn on the EMX console. • If you want to use the Gauss Meter start it up by double clicking the bootp desktop icon.
University of Minnesota College of Science and Engineering Characterization Facility Bruker D8 Discover User Manual (Version: 2012.08.08) Auto x, y, z Stage Reference. This product is covered by the terms of the Bruker Corporation standard warranty as in effect on the date of shipment and as reflected on the Bruker Order Acknowledgement and Quote. While a summary of the warranty statement is provided below, please refer to the Order Acknowledgement and/or Quote. Wondershare Dr Fone 4 Keygen Download.
Wait for a response indicating the bootp process has completed. Stow the bootp window. • Start the Win-EPR Acquisition software. The icon is on the desktop. • Turn on the magnet. Cavity Tuning • Click on the icon.
The Microwave Bridge Control window will pop up. • The bridge will be in the standby mode. Select Tune and set the microwave attenuation to 25 dB. • Wait a few seconds to allow the frequency to set. Adjust the frequency to locate the cavity dip and set it so that the frequency marker is at the bottom of the dip. This should be the point where the dip is the deepest. Kirk Rudy 527 Tabber Manual. • Select operate.
Set the microwave attenuation to 40 dB. The AFC meter should be in the green at this point. If not adjust the frequency until it is. If the frequency is close you can use the fine tune button. • Adjust the diode bias so that the meter is centered in the green.
• Lower the microwave attenuation by 10 dB. If the diode current meter deflects all the way to one side or the other increase the attenuation in 1 dB increments until the meter is on scale. • Adjust the iris up or down so that the diode current meter is centered.
Make sure the AFC stays in the green throughout the process. If not adjust the frequency to bring it back. • Repeat steps 5 and 6 until you have reached 10 dB and the diode meter is centered. • Adjust the phase so as to maximize the diode current.